INO
CNR
vai_a_storia   vai_a_organizzazione   vai_a_sedi   vai_a_personale   Area Riservata
    English English Version  
 
 

The gas absorption cell for GIANO-TNG

  Articoli su Riviste JCR/ISI  (anno 2006)

Autori:  D'Amato F., Oliva E., Origlia L

Affiliazione Autori:  CNR – Istituto Nazionale di Ottica Applicata, Largo. E. Fermi 6, 50125 Firenze, Italy; INAF – Osservatorio di Arcetri, Largo E. Fermi 5, 50125 Firenze, Italy; Telescopio Nazionale Galileo, calle A. de Abreu 70/1, 38700 S.Cruz de La Palma, Spain; INAF – Osservatorio di Bologna, via Ranzani 1, 40127 Bologna, Italy

Riassunto:  GIANO is an high resolution cross-dispersed spectrometer operating at near IR wavelengths (0.9-2.5 microns). One of its primary aims is to measure radial velocities of cool stars with an accuracy of a few m/sec. To this purpose, the spectrometer requires a gas absorption cell which should ideally produce several hundreds of measurable lines distributed over the whole wavelengths range. We show that this can be conveniently achived using a combination of halogen hydrates, namely HCl, HBr and HI. We present and discuss the design and development of such a cell.

Rivista/Giornale:  
Volume n.:        Pagine da: 62695E-1  a: 62695E-5
DOI: 10.1117/12.670375


Riferimenti visionabili in IsiWeb of Knowledge: (solo per sottoscrittori)
Per visualizzare la scheda dell'articolo su IsiWeb: Clicca qui

INO © Istituto Nazionale di Ottica - Largo Fermi 6, 50125 Firenze | Tel. 05523081 Fax 0552337755 - P.IVA 02118311006     P.E.C.    Info